废气、环境空气、公共场所空气和作业场所空气中氯乙烯检测填充柱2m*3mm
废气、环境空气、公共场所空气和作业场所空气中氯乙烯检测填充柱2m*3mm 详细信息:
名称:不锈钢柱
规格:2m*3mm*4mm,60-80目
型号:氯乙烯填充柱
编号:20230301001
zui高温度:270°C
浩瀚色谱(山东)应用技术开发有限公司建立工作场所、环境空气、废气中的氯乙烯的热解吸.气相色谱测定方法。方法 Carboxen1000吸附管吸附工作场所、环境空气中氯乙烯、二氯乙烷经热解吸.气相色谱进行定性及定量分析。结果标准曲线相关系数达到0.9998,检出限30ng/样品,zui低检出浓度0.02mg/m3(以采集1.5L空气计),解吸效率91.3%-94.5%,精密度0.9%-2.7%,吸附管净化好后和样品采集后用黄铜螺帽密封放入密封的塑料袋中常温可保存一周。结论:本方法可用于工作场所、环境空气、废气、室内空气中氯乙烯的检测。
废气、环境空气、公共场所空气和作业场所空气中氯乙烯检测填充柱2m*3mm 测试谱图:




2 * 3 mm filled column for vinyl chloride detection in waste gas, ambient air, public place air and workplace air
2 m * 3 mm filled column for vinyl chloride detection in waste gas, ambient air, public place air and workplace air Details:
Name: stainless steel column
Specification: 2m * 3mm * 4mm, 60-80 mesh
Model: vinyl chloride filled column
No.: 20230301001
Zui high temperature: 270 ° C
Haohan Chromatography (Shandong) Applied Technology Development Co., Ltd. has established a method for the determination of vinyl chloride in workplace, ambient air and waste gas by thermal desorption and gas chromatography. Methods Carboxen1000 adsorption tube was used to adsorb vinyl chloride and dichloroethane in the workplace and ambient air, and then pyrolysis and gas chromatography were used for qualitative and quantitative analysis. Results The correlation coefficient of the standard curve reached 0.9998, the detection limit was 30ng/sample, the low detection concentration of zui was 0.02mg/m3 (calculated by collecting 1.5L air), the desorption efficiency was 91.3% - 94.5%, and the precision was 0.9% - 2.7%. After the adsorption tube was purified and the sample was collected, it was sealed with brass nuts and put into a sealed plastic bag for one week at room temperature. Conclusion: This method can be used for the determination of vinyl chloride in workplace, ambient air, exhaust gas and indoor air.
Test spectrogram of 2 m * 3 mm filled column for vinyl chloride detection in exhaust gas, ambient air, public place air and workplace air: