气相色谱氩、氪、氙、氡分离5A分子筛填充柱
气相色谱氩、氪、氙、氡分离5A分子筛填充柱 详细信息:
名称:5?填充柱
规格:5mm(OD)*4mm(ID)*2m,40-60Mesh
使用温度:320°C
应用:GB/T 14599-1993 高纯氧
核bao炸产生多种稀有气体核素,这些核素可作为核泄露监测对象。气相色谱是组分分离的重要手段,在核爆炸监测取样中有重要应用。
浩瀚色谱(山东)应用技术开发有限公司以5A分子筛作为固定相,高纯氮作为流动相,测量了氩、氪、氙、氡在气相色谱中的分离效果。结果表明,在相同色谱条件下,氩、氪分离度小于氪、氙分离度;在柱流量35.6mL/min、柱温353K条件下,气相色谱氩、氪,氪、氙,氙、氡分离度均大于2.25。
气相色谱氩、氪、氙、氡分离5A分子筛填充柱 测试谱图:










Gas Chromatography Argon, Krypton, Xenon, Radon Separation 5A Molecular Sieve Packed Column
Gas Chromatography Argon, Krypton, Xenon, Radon Separation 5A Molecular Sieve Packed Column Details:
Name: 5? Packed column
Specifications: 5mm(OD)*4mm(ID)*2m, 40-60Mesh
Operating temperature: 320°C
Application: GB/T 14599-1993 High Purity Oxygen
Nuclear explosions produce a variety of rare gas nuclides, which can be used as nuclear leakage monitoring objects. Gas chromatography is an important means of component separation, and has important applications in nuclear explosion monitoring and sampling.
Haohan Chromatography (Shandong) Application Technology Development Co., Ltd. used 5A molecular sieve as the stationary phase and high-purity nitrogen as the mobile phase to measure the separation effect of argon, krypton, xenon and radon in gas chromatography. The results show that under the same chromatographic conditions, the separation degree of argon and krypton is lower than that of krypton and xenon; under the conditions of column flow rate of 35.6 mL/min and column temperature of 353 K, the separation degree of argon, krypton, krypton, xenon, xenon and radon of gas chromatography All are greater than 2.25.
Gas chromatography argon, krypton, xenon, radon separation 5A molecular sieve packed column Test spectrum: