model
型 号
|
hp-80
|
hp-150
|
hp-225
|
hp-800
|
|
internal dim. w×h×d
內部尺寸(cm)
|
40×50×40
|
50x60x50
|
50×75×60
|
100×100×80
|
|
external dim. w×h×d
外部尺寸(cm)
|
107×150×78
|
122×170×93
|
122×175×113
|
162×190×113
|
|
temp. range
温度范围
|
-70℃~+100℃ (+150℃)(a: 0℃ b:-20℃ c:-40℃ d:-60℃ e:-70℃)
|
|
relative humidity range
湿度范围
|
20%~98% r.h.
|
|
indication resolution
温湿度解析精度
|
0.1℃; 0.1% r.h. / 0.01℃; 0.1% r.h.
|
|
temp.&humi. constancy
温湿度控制精度
|
±0.5℃ ; ± 2.5% r.h.
|
|
temp.&humi. uniformity
温湿度分布精度
|
-40℃~100℃ (±1℃);-40℃↓~100℃ ↑ (±1.5℃);±3%rh
|
|
heat-up time (min)
升温时间
|
about 30 to 50
|
|
pull-down time (min)
降温时间
|
about 45 to 120
|
|
exterior
外部材质
|
镀锌钢板 + coating
|
|
interior
內部材质
|
sus#304 stainless steel plate
|
|
insulation
断热材料
|
pu & fiberglass wool
|
|
circulation
循环装置
|
sirocco fan
|
|
heating
加热器
|
sus#304 stainless steel high – speed heater
|
|
humidifying
加湿装置
|
surface evaporating system
|
|
cooling
冷却系统
|
airtight compressor + evaporator fins + air-cooling condenser
|
|
dehumidifying
除湿装置
|
adp critical dewpoint cooling / dehumidifying method
|
|
control
控制方式
|
digital electronic indicators. programmable computer controller + s.s.r
|
|
other components
其他配备
|
vacuum glass inspection window, cable port(50mm), control status indicator lights, chamber lamps, experiment separation shelves (freely adjustable 2 pcs.)
|
|
safety devices
安全裝置
|
overheating prevention circuit-breaker, compressor overload protection, control system overload prevention, humidifying system overheatingdrying out prevention, breakdown indicator light.
|
|
power source
电 源
|
ac 1ψ / 3 ψ, 220v ±10%, 50/60 hz
|