产品技术参数
Defi-B
运行方式: 异步,外部除颤治疗
电 极 板: 内含组合式儿童电极
能量等级: 20/50/100/160/250 /360焦耳(50Ω)
电池容量: 360焦耳放电45次(当显示电池用完后,还有10次放电能
量的储存
输 出: 抗空载运行和短路
充电时间: 充电至100焦耳约2秒钟,充电至360焦耳约7秒钟
能量供应: 蓄电池(PRIMEDICTM )
蓄 电 池: 14.4伏/1.4安培
外形尺寸: 48*40*12cm
重 量: 8.5Kg
电源电压: 200-240V 50/60Hz
Product technical parameters
Defi-B
Operation mode: asynchronous, external defibrillation therapy
Electrode plate: contains combined children's electrode
Energy level: 20 / 50 / 100 / 160 / 250 / 360 joules (50 Ω)
Battery capacity: 360 Joule discharge 45 times (when the display battery is
used up, there are 10 discharge times
Storage of quantity Output: resistance to no load operation and short circuit
Charging time: about 2 seconds to 100 joules and 7 seconds to 360 joules
Energy supply: battery (primedictm)
Battery: 14.4 V / 1.4 A
Dimension: 48 * 40 * 12cm
Weight: 8.5Kg
Power supply voltage: 200-240v 50 / 60Hz