现货MI-414SeriesMicro-Com…

现货MI-414SeriesMicro-CombinationpHProbe(16gaugeneedle)
产品简介
详细介绍
  • 参考报价:电议 产地:美国 品牌:microelectrodes 型号:MI-414 更新时间:2017/12/29

f659cb_9f8d35569274422da054d81358335b5c_mv2.jpg

SPECIFICATIONS


TOTAL LENGTH                             

15.7 cm

LEAD LENGTH

1 m  

BODY (Outer Diameter)

7.6 mm

TIP (Outer Diameter)

pH: 0.75 mm

Ref + pH: 1.3 mm 

RESPONSE TIME:

5 to 15 seconds  

(stirred solution)

DEPTH of IMMERSION:

0.5 mm to 1 mm  

SENSITIVITY:

pH 0 to 14  

TEMPERATURE RANGE:

-5°C to 100° C  

REFERENCE ELECTRODES

TYPE: Solution & Electrode

 

3M KCI & Ag-AgCI


CALIBRATION


Carefully unwind the tape and remove the probe from the protective glass tube.  The white translucent sleeve, which covers the hole, should be moved down the glass barrel slightly to ventilate the reference chamber and replaced after each use. 

 

The use of a separate Reference Electrode is not required.

 

The probe can be held by hand when measuring pH.  It has been standardized in two (2) pH buffers.  Follow the procedure recommended by the manufacturer of your pH meter for calibrating your pH meter with our microelectrode.

 

OPTIMUM RESPONSE TIME will be obtained after the probe has been exercised in two (2) buffer solutions. 

 

  • Fill one beaker with a pH 4 buffer or equivalent and a second beaker with a pH 7 buffer or equivalent. 

  • Hold the microelectrode and touch the pH 4 buffer surface allowing 5 seconds for equilibration, and then touch the pH 7 buffer surface in the same manner.  Do this several times. 

  • It is not necessary to rinse the probe because the carry-over is negligible.


CLEANING


Using any electrode in solutions containing protein requires the electrode be rinsed with an enzyme cleaning solution.

 

After each use, we recommend cleaning our electrodes with Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning solution by submerging the electrodes for a couple of minutes in order to remove all protein from the glass and reference junction.

 

This will prolong the useful life of the electrodes.


STORAGE


Always clean the microelectrode before storing:

 

  • Long-term (over 2 weeks): Return the probe to its original container and prepare it in the same condition in which you received it.  Usually this means simply moistening the sponge located in the bottom of the protective glass tube with pH 4 buffer.

  • Short-term: The probe can be left in an acid pH buffer solution (pH 4.01).