CL510A-1_CL510A-10欧米茄RTD模拟器OMEGA
源一条或多条RTD曲线
高精度
易于阅读的背光显示屏
直接在温度下校准(°C或°F)
与所有RTD工艺仪表兼容
CL510A-1_CL510A-10欧米茄RTD模拟器OMEGA
SPECIFICATIONS (Unless otherwise indicated, all specifications are rated from a nominal 23°C (73°F), 70% RH)
Accuracy: ±(0.015% of Rdg + 0.05 Ω)
Temperature Drift: ±0.05 Ω/°C
Operating Temperature Range: -25 to 60°C (-10 to 140°F)
Relative Humidity Range, 10% ≤RH:
≤90%: 0 to 35°C (32 to 95°F), non-condensing
≤70%: 35 to 60°C (95 to 140°F), non-condensing
Size: 126 L x 69 W x 45 mm H (4.96 x 2.73 x 1.79")
Weight: 0.24 kg (8.4 oz) (including batteries)
Batteries: 4 "AA" alkaline 1.5V (included)
Battery Life: 50 hr
Low Battery: Low battery indication with nominal 1 hour of operation left
Protection Against Misconnection: Over-voltage protection to 60 Vdc (rated for 30 sec)
Display: High contrast graphic liquid crystal display, LED backlighting for use in low lit areas
Source
Accuracy: From 1 to 10.2 mA
External Excitation Current: ±(0.015% of rdg + 0.05 Ω) Below 1 mA of External
Excitation Current: ±(0.015% of rdg + 0.01 mV/mA excitation current + 0.05 Ω)
Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω
Allowable Excitation Current Range, Steady or Pulsed/Intermittent:
<410 Ω: 10.2 mA max
410 to 001 Ω: 1 mA max Pulsed Excitation Current
Compatibility: DC to 0.01 second pulse width
CL510A-1_CL510A-10欧米茄RTD模拟器OMEGA