CCT-22-0/400C_CCT-23-0/10…

CCT-22-0/400C_CCT-23-0/1000C_CCT-24-0/200C_CCT-24
产品简介
详细介绍
  • 参考报价:暂无 产地:美国 品牌:欧米茄 型号:CCT-22-0/400C 添加时间:2019/8/20 8:12:05

CCT-22-0/400C_CCT-23-0/1000C_CCT-24-0/200C_CCT-24-0/300C信号调节器OMEGA

Voltage, Current, Frequency, Resistance, Potentiometer, Thermocouple, RTD and Strain Gage Input Modules

Integrator Modules Convert Voltage or Current to Frequency

Loop Powered Galvanic Isolator Provides 2 kVeff Isolation Between Input and Output

Compatible with Standard 35 mm DIN Rail


The CCT series DIN rail signal conditioners are designed to accept a broad range of input signals, such as ac and dc voltage and current, frequency, temperature (thermocouple and RTD), and process transducers, and provide standard process outputs of either 0/4 to 20 mA, or 0 to 10 Vdc. The CCT series features a modern housing design, that is easily mounted on standard DIN rails. Connections are safely and securely made, with input and output connections on the opposite sides of the module.


Internal Design

The CCT series is designed using an internal plug-in three module system that provides flexibility in selecting and changing the power supply, input signal and output signal. The units are supplied with 110 Vac power standard, Consult sales for 220 Vac, and 24 Vdc power.


Isolation

The three internal modules in each signal conditioner (signal input, signal output and power supply) are isolated up to 2 kVeff.


Outputs

Each CCT series signal conditioner is available with current and voltage output (only one may be used at a time). Available output types include 4 to 20 mA or 0 to 20 mA (jumper-selectable) and 0 to 10 Vdc.


Standard outputs are linear and proportional to the signal input. Thermocouple input modules feature special circuitry to linearize the output to the actual temperature, and not to the non-linear signal produced by thermocouple sensors.


Specifications

Power: 110 Vac (standard); 220 Vac or 24 Vdc (optional) (±10% tolerance on all power supplies)

Output: 4 to 20 mA and 0 to 10 Vdc standard; 0 to 20 mA via jumper select

Accuracy: ≤0.2% or ≤0.3,%, depending on the model

Response Time: ≤250 ms

Isolation: 2 kVeff

Ripple: ≤0.5%

Bandpass: 1.5 Hz (-3 dB)

Operating Ambient: -10 to 60°C (14 to 140°F)

Storage Ambient: -30 to 80°C (-22 to 176°F)

Temperature Coefficient: ≤0.015%/°C

Power Consumption: 1.5 VA max

Enclosure: Polycarbonate, RAL 7032; UL 94 V-1, IP40 protection

Terminal Housing: Polycarbonate, UL 94-2, IP20 protection

Max Wire Size: 4 mm2 (14 AWG)

Dimensions: 75 H x 45 W x 110 mm D (2.95" x 1.77" x 4.33")

Weight: 270 g (0.6 lb)

Mounting: DIN rail 46277 and EN 50022 (35 x 7.5 mm)