产品介绍
美国热电酶标仪Multiskan FC,进口优秀品牌酶标仪,我公司热销活动进行中!!!
Thermo Fisher Scientific Inc. (NYSE: TMO) is the world leader in serving science, with revenues of $17 billion and more than 50,000 employees in 50 countries. Our mission is to enable our customers to make the world healthier, cleaner and safer. We help our customers accelerate life sciences research, solve complex analytical challenges, improve patient diagnostics and increase laboratory productivity. Through our premier brands – Thermo Scientific, Applied Biosystems, Invitrogen, Fisher Scientific and Unity Lab Services – we offer an unmatched combination of innovative technologies, purchasing convenience and comprehensive support.
美国热电酶标仪Multiskan FC技术参数:
【1】光源:石英卤素灯
【2】波长范围:340-850nm
【3】滤光片:8位滤光片,标配3块滤光片:405nm,450nm和620nm,其它滤光片可选
【4】滤光片半带宽:3-9nm
【5】读数范围:0-6Abs
【6】线性范围(405nm):0-3Abs,96孔板,快速测量模式
0-4Abs,96孔板,标准测量模式
【7】分辨率:0.001Abs
【8】准确性(405nm):±1%(0-3Abs),±2%(3-4Abs)
【9】精确性(405nm):CV≤0.2%(0-3Abs),标准测量模式
【10】测量速度:6s,96孔板,快速测量模式;12s,96孔板,标准测量模式
11s,384孔板,快速测量模式;33s,384孔板,标准测量模式
【11】孵育器(选配):温控范围:室温+4℃至50℃
【12】振荡器:线性震荡,三档速度可选
【13】机械臂兼容:可以
【14】按键和显示:高分辨彩色显示屏,分辨率为480X272像素
【15】用户界面:内置软件和Skanllt软件(通过电脑控制)
【16】内存:仪器内可存储100个测量程序和100组测量结果(96孔板)
【17】外接打印机类型:HP PCL5
【18】通讯接口:USB用于计算机连接
USB外接U盘用于结果的输出和保存
USB外接打印机接口
【19】电源:100-240V(50/60Hz)
【20】功率:最大100W,待机8W
【21】外部尺寸:210mm(H)X290mm(W)X400mm(D)
【22】重量:8.5kg
美国热电酶标仪Multiskan FC如何设置震荡
1. 使用向下箭头键选择主菜单中的“振荡”行,然后按 OK(确定)按钮。
2. 在“模式”项目上按 OK(确定)按钮并使用向下箭头键进行选择,例如选择 Before measurement(测量前)作为振荡模式,然后按 OK(确定)按钮。
3. 使用向下箭头键选择“速度”项目,然后按 OK(确定)按钮。
4. 使用向下箭头键选择,例如选择 Fast(快)作为振荡速度,然后按 OK(确定)按钮。
5. 使用向下箭头键选择“时间 (hh:mm:ss)”项目,然后按 OK(确定)按钮。
6. 使用数字键选择,例如选择 20 秒 (= 00:00:20) 作为振荡时间,然后按 OK(确定)按钮。
7. 按 F1 键接受选择并返回到主菜单。