产品介绍
特性:
Microsaic Systems公司推出的全球首款基于芯片的质谱仪Microsaic 3500 MiD
英国新一代质谱仪技术开发商——Microsaic Systems公司在德国慕尼黑举办的分析、生化及实验室技术展(Analytica
2012)期间展出全球首款基于芯片的质谱仪Microsaic 3500 MiD。
据悉,该款质谱仪由三大制药公司合作开发,较常规质谱仪小90%。Microsaic Systems公司已通过将关键的微型元件与专利的芯片技术相结合,成功地改变了质谱分析法。其核心技术是传统质谱组件的芯片级版本。3500 MiD也是目前市面上唯一一款不需要外部真空泵的质谱系统。
另悉,该质谱仪的设计旨在与现有的高效液相色谱设备集成,无需额外封装及基础设施。3500 MiD也可与Prep LC、Flash LC和SFC等其它分离方法结合。
简介:
The Microsaic 3500 MiD is the world’s smallest fully integrated MS system. The 3500 MiD is the size of a standard desktop computer and with very few additional infrastructure requirements it is smaller, lighter, more energy efficient and cheaper to run than conventional MS systems.
The 3500 MiD is designed to readily integrate with existing HPLC equipment with no extra footprint extending the capability of your HPLC system with significant cost savings compared to a conventional MS system.
Historically, MS has only been used for off-line analysis but the 3500 MiD is so compact it can be used for on-line reaction monitoring in a fume-hood. Monitoring batch reactions in real time provides the chemist with vital reaction data which can be acted upon to maximise yield and/or minimize impurities. The 3500 MiD is also ideal for flow chemistry systems where MS data can be used for faster reaction optimisation.
The system is portable with modular plug and play components for maximum application flexibility.
特点:
Far smaller than conventional mass spectrometers, with fewer infrastructure requirements (e.g. sound- proofing, floor pump, UPS, gas generator) the Microsaic 3500 MiD is a versatile, portable solution designed for bench chemists.
Designed to integrate with existing equipment with no extra footprint.
The 3500 MiD is a fully integrated MS system. The core technologies are chip-scale versions of traditional mass spec components which can be interchanged rapidly by the user. This modular approach allows users to maintain the system without the need for expensive service contracts and time-consuming call-outs.
Microspray ionisation source
Micro-engineered vacuum interface
Chip-based quadrupole mass analyzer
‘Plug and play’ chips make the 3500 MiD configurable, flexible and future proof.
The 3500 MiD runs Masscape; an intuitive, open-access software for method development and LC-MS.
Masscape may be used off-line as a data analysis solution. Masscape files are in a variety of formats including NetCDF, compatible with software from major LC companies.
10% of the running costs of conventional MS systems. Big savings on costly infrastructure such as gas generators, air-con and UPS.
‘Plug and play’ components allow users to maintain the system. Zero call outs.
Remote control via LAN connection removes the need for a monitor or keyboard, further minimizing footprint: ideal for fume-hood deployment.
PHYSICAL REQUIREMENTS |
|
---|---|
Dimensions |
60 cm * 35 cm * 18.5 cm (L) * (W) * (H) |
Weight |
27 kg |
HARDWARE |
|
Ion Source |
spraychip® |
Ionization Modes |
Positive and negative ESI |
Sample Inlet |
Compatible with standard LC capillarys and unions e.g. 1/16" tubing |
Direct Flow Rate |
0.2µL/min - 1µL/min |
Split Flow Rate |
1µL/min – 1.8ml/min |
Mass Analyzer |
ionchip® quadrupole mass spectrometer |
Mass Range |
80-800 m/z with ionchip®150 |
Mass Accuracy |
±0.3 m/z |
Mass Resolution |
0.7 FWHM |
Sensitivity |
10pg with a S/N of 10:1 rms in SIM mode using a 1µL injection of Reserpine at 150µL/min |
Spectrum Display |
Continuum (profile) and Centroid (stick) |
Vacuum Interface |
vac-chip™ microengineered API |
Detector |
Electron multiplier |
Dynamic Range |
3 orders of magnitude |
Scan Modes |
Full scan or 8 m/z channels (SIM) |
Scan Rate |
Up to 3,600 pts/s |
Pumping System |
Integrated maintenance-free pumps. |
Computer |
Integrated on-board computer |
Operating System |
Linux Open Source |
Connectivity |
Ethernet, 3 x USB |
ENVIRONMENTAL REQUIREMENTS |
|
Operating |
10-35 °C with 40%~80% relative humidity noncondensing. |
Storage |
-20-70 °C with 0%~90% relative humidity noncondensing |
REGULATORY COMPLIANCE |
|
---|---|
Safety |
IEC 61010-1: 2001 |
MASSCAPE™ SOFTWARE |
|
The Masscape™ software application gives you full control of the data acquisition and analysis of your chemical samples. |
|
Data Capture |
Mass spectra (profile and centroid) selected ion monitoring modes (SIM) |
Data Analysis |
Peak integration |
Chromatogram |
Retention time, peak height and area (integrated chromatogram) |
Mass Spectrum |
m/z and intensity profile or centroid |
Data Input |
Real-time analog input of UV trace and display in Masscape |
Data Output |
Real-time analog output of TIC or SIM trace |
Data Export |
NetCDF, Excel, CSV |
Configuration |
May be controlled remotely over LAN. External trigger allowing automatic analysis of samples via a method table. |
UTILITY REQUIREMENTS |
|
Power |
200 W (operating) 300 W (peak) |
Nitrogen Gas |
Pressure 2- 6 bar (29 – 87 psi) |
Pump Exhaust |
Exhaust connects to 2 * 8 mm OD push fit bulkhead connections |
规格说明:
PHYSICAL REQUIREMENTS |
|
---|---|
Dimensions |
60 cm * 35 cm * 18.5 cm (L) * (W) * (H) |
Weight |
27 kg |
HARDWARE |
|
Ion Source |
spraychip® |
Ionization Modes |
Positive and negative ESI |
Sample Inlet |
Compatible with standard LC capillarys and unions e.g. 1/16" tubing |
Direct Flow Rate |
0.2µL/min - 1µL/min |
Split Flow Rate |
1µL/min – 1.8ml/min |
Mass Analyzer |
ionchip® quadrupole mass spectrometer |
Mass Range |
80-800 m/z with ionchip®150 |
Mass Accuracy |
±0.3 m/z |
Mass Resolution |
0.7 FWHM |
Sensitivity |
10pg with a S/N of 10:1 rms in SIM mode using a 1µL injection of Reserpine at 150µL/min |
Spectrum Display |
Continuum (profile) and Centroid (stick) |
Vacuum Interface |
vac-chip™ microengineered API |
Detector |
Electron multiplier |
Dynamic Range |
3 orders of magnitude |
Scan Modes |
Full scan or 8 m/z channels (SIM) |
Scan Rate |
Up to 3,600 pts/s |
Pumping System |
Integrated maintenance-free pumps. |
Computer |
Integrated on-board computer |
Operating System |
Linux Open Source |
Connectivity |
Ethernet, 3 x USB |
ENVIRONMENTAL REQUIREMENTS |
|
Operating |
10-35 °C with 40%~80% relative humidity noncondensing. |
Storage |
-20-70 °C with 0%~90% relative humidity noncondensing |
REGULATORY COMPLIANCE |
|
---|---|
Safety |
IEC 61010-1: 2001 |
MASSCAPE™ SOFTWARE |
|
The Masscape™ software application gives you full control of the data acquisition and analysis of your chemical samples. |
|
Data Capture |
Mass spectra (profile and centroid) selected ion monitoring modes (SIM) |
Data Analysis |
Peak integration |
Chromatogram |
Retention time, peak height and area (integrated chromatogram) |
Mass Spectrum |
m/z and intensity profile or centroid |
Data Input |
Real-time analog input of UV trace and display in Masscape |
Data Output |
Real-time analog output of TIC or SIM trace |
Data Export |
NetCDF, Excel, CSV |
Configuration |
May be controlled remotely over LAN. External trigger allowing automatic analysis of samples via a method table. |
UTILITY REQUIREMENTS |
|
Power |
200 W (operating) 300 W (peak) |
Nitrogen Gas |
Pressure 2- 6 bar (29 – 87 psi) |
Pump Exhaust |
Exhaust connects to 2 * 8 mm OD push fit bulkhead connections |
世联博研北京科技有限公司是英国Microsaic Systems在中国授权代理商,咨询、索取产品资料请致电010-67529703