96T 大鼠催乳素(PRL)ELISA试剂盒说明书

来宝网 2012/3/3点击1430次

 

 Rat prolactin(PRL)ELISA Kit

Catalog No. CSB-E06881r

(96 tests)

This immunoassay kit allows for the in vitro rapid detection of rat PRL concentrations in serum, plasma.

Expiration date six months from the date of manufacture

FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

厦门慧嘉生物长期经营ELISA试剂盒及Santa/Abcam抗体、Prospec细胞因子、Sigma/Amresco/Qiagen、Axygen耗材等生物试剂产品。诚信经营,为客户提供“最高质量的产品”和“最优质的服务”:18906011628  0592-6020891 QQ: 1193952234 1048735792 或登陆http://www.biohj.com(向客服人员索取原版说明书)。欢迎广告老师来询!咨询电话

 

PRINCIPLE OF THE ASSAY

The microtiter plate provided in this kit has been pre-coated with a goat-anti-rabbit antibody. Sta-n-dards or samples are then added to the appropriate microtiter plate wells with a HRP-conjugated PRL a-n-d antibody preparation specific for PRL a-n-d incubated. Then substrate solutions are added to each well. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution a-n-d the color change is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of PRL in the samples is then determined by comparing the

O.D. of the samples to the sta-n-dard curve.

DETECTION RANGE

The sta-n-dard curve concentrations used for the ELISA’s were 50ng/ml, 15ng/ml, 2.5ng/ml, 0.5ng/ml,0.2 ng/ml.

SPECIFICITY

This assay recognizes rat PRL. No significant cross-reactivity or interference was observed.

MATERIALS PROVIDED

Reagent

Quantity

Assay plate

1

Standards (S0-S5)

6x 0.5 ml

Antibody HRP-conjugated

1 x 6 ml 1 x 6 ml

Wash Buffer

1 x 15 ml (20×concentrate)

Substrate A

1 x 7 ml

Substrate B

1 x 7 ml

Stop Solution

1 x 7 ml

 

Standard

S0

S1

S2

S3

S4

S5

Concentration(ng/ml)

0

0.2

0.5

2.5

15

50

STORAGE

1          Unopened test kits should be stored at 2-8?C upon receipt a-n-d the microtiter plate should be kept in a sealed bag to minimize exposure to damp air. The test kit may be used throughout the expiration date of the kit. Refer to the package label for the expiration date.

2          A microtiter plate reader with a ba-n-dwidth of 10 nm or less a-n-d an optical density range of 0-3 OD or greater at 450nm wavelength is acceptable for use in absorbance measurement.

 

TECHNICAL HINTS

1          Bring all reagents a-n-d plate to room temperature for at least 30 minutes before use. Unused wells need store at 2-8a-n-d avoid sunlight.

2          Wash Buffer If crystals have formed in the concentrate, warm to room temperature a-n-d mix gently until the crystals have completely dissolved. Dilute 15 ml of Wash Buffer Concentrate into deionized or distilled water to prepare 300 ml of Wash Buffer.

3          To avoid cross-contamination, change pipette tips between additions of each sta-n-dard level, between sample additions, a-n-d between reagent additions. Also, use separate reservoirs for each reagent.

4          When using an automated plate washer, adding a 30 second soak period following the addition of wash buffer, a-n-d/or rotating the plate 180 degrees between wash steps may improve assay precision.

5          Substrate Solution should remain colorless until added to the plate. Keep Substrate Solution protected from light. Substrate Solution should change from colorless to gradations of blue.

6          Stop Solution should be added to the plate in the same order as the Substrate Solution. The color developed in the wells will turn from blue to yellow upon addition of the Stop Solution. Wells that are green in color indicate that the Stop Solution has not mixed thoroughly with the Substrate Solution.

 

Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye, ha-n-d, face, a-n-d clothing protection when using this material.

OTHER SUPPLIES REQUIRED

1           Microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.

2           Pipettes a-n-d pipette tips.

3           Deionized or distilled water.

4           Squirt bottle, manifold dispenser, or automated microplate washer.

 

SAMPLE COLLECTION A-N-D STORAGE

Serum Use a serum separator tube (SST) a-n-d allow samples to clot for 30 minutes before centrifugation for 15 minutes at 1000 x g. Remove serum a-n-d assay immediately or aliquot a-n-d store samples at -20°C. Avoid repeated freeze-thaw cycles.

Plasma Collect plasma using citrate, EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 x g within 30 minutes of collection. Assay immediately or aliquot a-n-d store samples at -20° C. Avoid repeated freeze-thaw cycles.

 

Note: Grossly hemolyzed samples are not suitable for use in this assay.

ASSAY PROCEDURE

Bring all reagents a-n-d samples to room temperature before use. It is recommended that all samples, sta-n-dards, a-n-d controls be assayed in duplicate.

1          Set a Blank well without any solution. Add 50μl of Sta-n-dard or Sample per well. Sta-n-dard need test in duplicate.

2          Add 50μl of HRP-conjugate to each well (not to Blank well), then 50μl Antibody to each well. Mix well a-n-d then incubate for 1 hour at 37°C.

3          Fill each well with Wash Buffer (about 250μl), stay for 10 seconds a-n-d Spinning. Repeat the process for a total of three washes. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate a-n-d blot it against clean paper towels.

4          Add 50μl of Substrate A a-n-d Substrate B to each well, mix well. Incubate for 15 minutes at 37°C. Keeping the plate away from drafts a-n-d other temperature fluctuations in the dark.

5          Add 50μl of Stop Solution to each well.

6          Determine the optical density of each well within 10 minutes, using a microplate reader set to 450 nm.

 

CALCULATION OF RESULTS

Average the duplicate readings for each sta-n-dard, Blank, a-n-d sample a-n-d subtract the optical density of Blank. Create a sta-n-dard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a sta-n-dard curve by plotting the mean absorbance for each sta-n-dard on the x-axis against the concentration on the y-axis a-n-d draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the PRL concentrations versus the log of the O.D. a-n-d the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the sta-n-dard curve must be multiplied by the dilution factor.

LIMITATIONS OF THE PROCEDURE

1           The kit should not be used beyond the expiration date on the kit label.

2           Do not mix or substitute reagents with those from other lots or sources.

3           Any variation in Sta-n-dard Diluent, operator, pipetting technique, washing technique, incubation time or temperature, a-n-d kit age can cause variation in binding.

4           This assay is designed to eliminate interference by soluble receptors, binding proteins, a-n-d other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.