      Porcine Interleukin1β(IL-1β)

                            ELISA Kit

 

                      Catalog No. CSB-E06782p

 

                                    (96T)

 

?   This immunoassay kit allows for the in vitro quantitative determination of porcine

 

    IL-1β concentrations in serum, plasma.

 

?   Expiration date   six months from the date of manufacture

 

?   FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

 

                                       1


----------------------- Page 2-----------------------

PRINCIPLE OF THE ASSAY

 

The microtiter plate provided in this kit has been pre-coated with

 

an   antibody   specific   to   IL-1β.   Standards   or   samples   are   then

 

added to the appropriate microtiter plate wells with a Horseradish

 

Peroxidase (HRP)-conjugated monoclonal antibody preparation

 

specific for IL-1β and incubated. Then substrate solution A and B

 

are   added   to   each   well.   Only   those   wells   that   contain   IL-1β,

 

HRP-conjugated   antibody   will   exhibit   a   change   in   color.   The

 

enzyme-substrate   reaction   is   terminated   by   the   addition   of   a

 

sulphuric     acid   solution   and    the   color   change     is  measured

 

spectrophotometrically at a wavelength of 450 nm ± 2 nm. The

 

concentration   of   IL-1β in   the   samples   is   then   determined   by

 

comparing the O.D. of the samples to the standard curve.

 

DETECTION RANGE

 

The   standard   curve   concentrations   used   for   the   ELISA’s   were

 

250pg/ml,        125pg/ml,       62.5pg/ml,       31.2pg/ml,       12.5pg/ml,

 

6.25pg/ml.

 

SPECIFICITY

 

This   assay   recognizes   recombinant   and   natural   porcine   IL-1β.

 

No significant cross-reactivity or interference was observed.

 

                                       2


----------------------- Page 3-----------------------

SENSITIVITY

 

The minimum detectable dose of porcine IL-1β is typically less

 

than 3.125 pg/ml. The sensitivity of this assay, or Lower Limit of

 

Detection   (LLD)   was   defined   as   the   lowest   concentration   that

 

could be differentiated from zero.

 

MATERIALS PROVIDED

 

Reagent                                               Quantity

Assay plate                                                1

Standard(S  -S  )                                          6

              1   6

 

HRP-conjugate                                          1 x 6 ml

                                                      1 x 15 ml

Wash Buffer

                                                 (20×concentrate)

Substrate A                                            1 x 6 ml

Substrate B                                            1 x 6 ml

Stop Solution                                          1x 6 ml

 

   Standard           S1        S2        S3        S4        S5        S6

Concentration

                     6.25      12.5      31.2      62.5      125       250

    (pg/ml)

 

STORAGE

 

1.  Unopened   test   kits   should   be   stored   at   2-8?C   upon   receipt

 

    and the microtiter plate should be kept in a sealed bag. The

 

    test kit may be used throughout the expiration date of the kit,

 

    provided     it  is  stored  as   prescribed    above.     Refer   to  the

 

    package label for the expiration date.

 

                                      3


----------------------- Page 4-----------------------

2.   Opened test plate should be stored at 2-8?C in the aluminum

 

   foil bag with desiccants to minimize exposure to damp air. The

 

    kits will remain stable until the expiring date shown, provided it

 

    is stored as prescribed above.

 

3.  A  microtiter   plate   reader   with   a   bandwidth   of   10   nm   or   less

 

    and an optical density range of 0-3 OD or greater at 450nm

 

   wavelength         is   acceptable       for    use     in    absorbance

 

    measurement.

 

REAGENT PREPARATION

 

1.  Bring all reagents and plate to room temperature for at least

 

    30 minutes before use. Unused wells need store at 2-8°C and

 

    avoid sunlight.

 

2.  Wash   Buffer       If   crystals   have   formed   in   the   concentrate,

 

   warm   to   room   temperature   and   mix   gently   until   the   crystals

 

    have    completely     dissolved.    Dilute   15  ml   of  Wash     Buffer

 

    Concentrate into deionized or distilled water to prepare 300 ml

 

   of Wash Buffer.

 

OTHER SUPPLIES REQUIRED

 

?  Microplate reader capable of measuring absorbance at 450

 

    nm, with the correction wavelength set at 540 nm or 570 nm.

 

?  Pipettes and pipette tips.

 

                                      4


----------------------- Page 5-----------------------

?  Deionized or distilled water.

 

?  Squirt bottle, manifold dispenser, or automated microplate

 

    washer.

 

?  An incubator which can provide stable incubation conditions

 

    up to 37°C±0.5°C.

 

SAMPLE COLLECTION AND STORAGE

 

?   Serum        Use     a   serum     separator     tube   (SST)    and    allow

 

    samples   to   clot   for   30   minutes   before   centrifugation   for   15

 

     minutes at 1000 g. Remove serum and assay immediately or

 

    aliquot   and   store   samples   at  -20°C.   Centrifuge   the   sample

 

    again      after   thawing     before   the    assay.     Avoid    repeated

 

    freeze-thaw cycles.

 

?    Plasma      Collect plasma using citrate, EDTA, or heparin as

 

    an anticoagulant. Centrifuge for  15 minutes at 1000 g within

 

    30   minutes   of   collection.   Assay   immediately   or   aliquot   and

 

    store   samples   at   -20°C.   Centrifuge   the   sample   again   after

 

    thawing before the assay. Avoid repeated freeze-thaw cycles.

 

Note: Grossly hemolyzed samples are not suitable for use in this assay.

 

ASSAY PROCEDURE

 

Bring   all  reagents  and   samples   to  room   temperature   before  use.  It  is

recommended that all samples, standards, and controls be assayed in duplicate.

All   the   reagents   should   be   added   directly   to   the  liquid   level   in   the   well.   The

pipette should avoid contacting the inner wall of the well.

 

                                        5


----------------------- Page 6-----------------------

1.   Set a Blank well without any solution. Add 100μl of Standard

 

    or Sample per well. Standard need test in duplicate.

 

2.   Add 50μl of HRP-conjugate to each well (not to Blank well).

 

    Mix well and then incubate for 2 hours at 37°C.

 

3.   Complete   remove   the   liquid.   Then   fill   each   well   with  Wash

 

    Buffer   (about     200μl),    stay   for  10  seconds     and   Spinning.

 

    Repeat   the   process   for   a   total  of   three   washes.   Complete

 

    removal       of  liquid   at   each    step    is   essential    to   good

 

    performance.       After   the   last  wash,    remove    any    remaining

 

    Wash Buffer by aspirating or decanting. Invert the plate and

 

    blot it against clean paper towels.

 

4.   Add 50μl of Substrate A and 50μl of Substrate B to each

 

    well, mix well. Incubate for 15 minutes at 37°C. Keeping the

 

    plate away from drafts and other temperature fluctuations in

 

    the dark.

 

5.   Add 50μl of Stop Solution to each well. If color change does

 

    not appear uniform, gently tap the plate to ensure thorough

 

    mixing.

 

6.   Determine the optical density of each well within 10 minutes,

 

    using a microplate reader set to 450 nm.

 

CALCULATION OF RESULTS

 

Using   the   professional   soft   "Curve   Exert 1.3"   to   make   a   standard   curve   is

recommended, which can be downloaded from our web.

 

                                       6


----------------------- Page 7-----------------------

Average the duplicate readings for each standard, control, and

 

sample and subtract the average zero standard optical density.

 

Create   a   standard   curve   by   reducing   the   data   using   computer

 

software capable of generating a four parameter logistic (4-PL)

 

curve-fit. As an alternative, construct a standard curve by plotting

 

the   mean   absorbance   for   each  standard   on   the   y-axis   against

 

the concentration on the x-axis and draw a best fit curve through

 

the points on the graph. The data may be linearized by plotting

 

the log of the IL-1βconcentrations versus the log of the O.D. and

 

the best fit line can  be determined by regression analysis. This

 

procedure   will   produce   an  adequate   but   less   precise   fit   of   the

 

data. If samples have been diluted, the concentration read from

 

the standard curve must be multiplied by the dilution factor.

 

LIMITATIONS OF THE PROCEDURE

 

?  The kit should not be used beyond the expiration date on the

 

    kit label.

 

?  Do not mix or substitute reagents with those from other lots or

 

    sources.

 

?  If samples generate values higher than the highest standard,

 

    dilute the samples with the appropriate Diluent and repeat the

 

    assay.

 

                                       7


----------------------- Page 8-----------------------

?  Any      variation    in  operator,    pipetting    technique,     washing

 

    technique,   incubation   time   or  temperature,   and   kit   age   can

 

    cause variation in binding.

 

?  This   assay   is   designed   to   eliminate   interference   by   soluble

 

    receptors,     binding    proteins,  and     other   factors   present    in

 

    biological samples. Until all factors have been tested in the

 

    Immunoassay,         the   possibility   of   interference    cannot     be

 

    excluded.

 

TECHNICAL HINTS

 

?  Centrifuge vials before opening to collect contents.

 

?  When mixing or reconstituting protein solutions, always avoid

 

    foaming.

 

?  To   avoid   cross-contamination,   change   pipette  tips   between

 

    additions of each standard level, between sample additions,

 

    and between reagent additions. Also, use separate reservoirs

 

    for each reagent.

 

?  When using an automated plate washer, adding a 30 second

 

    soak     period   following   the   addition   of  wash    buffer,  and/or

 

    rotating    the  plate   180   degrees   between      wash    steps    may

 

    improve assay precision.

 

?  To ensure accurate results, proper adhesion of plate sealers

 

    during incubation steps is necessary.

 

                                       8


----------------------- Page 9-----------------------

?  Substrate Solution should remain colorless or light blue until

 

    added to the plate. Keep Substrate Solution protected from

 

    light. Substrate Solution should change from colorless or light

 

    blue to gradations of blue.

 

?  Stop Solution should be added to the plate in the same order

 

    as the Substrate Solution. The color developed in the wells

 

    will turn from blue to yellow upon addition of the Stop Solution.

 

    Wells that are green in color indicate that the Stop Solution

 

    has not mixed thoroughly with the Substrate Solution.

 

                                       9

 