       Mouse Endothelin 1                               ((ET-1))

                                                        ((  ))

 

                            ELISA Kit

 

                      Catalog No. CSB-E05145m

 

                                    (96T)

 

    This immunoassay kit allows for the in vitro quantitative determination of mouse

 

    ET-1   concentrations  in cell  culture  supernates,  serum,  plasma   and  other

 

    biological fluids.

 

    Expiration date   six months from the date of manufacture

 

    FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

 

                                       1


----------------------- Page 2-----------------------

INTRODUCTION

 

Endothelin-1(ET-1), a peptide of 21 amino acid residues, is the most potent

 

vasoconstrictive   substance   known.   Originally   isolated   from   porcine   aortic

 

endothelial     cells,  ET-1    is  now   known     to  be   one   of  a  family   of  three

 

mammalian vasoactive peptides that also includes Endothelin-2 (ET-2) and

 

Endothelin-3 (ET-3). These related peptides differ from ET-1 at two and six

 

amino   acid   residue   positions,   respectively.   A   fourth   peptide,   vasoactive

 

intestinal    contractor    (VIC),    is  sometimes        classified   as   rat  ET-2.    All

 

members   of   the   endothelin   family   contain   two   essential   disulfide   bridges

 

and six conserved amino acid residues at the C-terminus. Additionally, all of

 

the      endothelin       family     members        are     synthesized        initially   as

 

prepropolypeptides of approximately 200 amino acid residues encoded by

 

separate       genes.      These      are    proteolytically     cleaved      to    produce

 

biologically-inactive      propolypeptides       of   approximately       40   amino     acid

 

residues   termed   “big  endothelins”.   Big   ET-1   is   cleaved   by   the   proteolytic

 

action     of   a   membrane-bound          metalloprotease       [endothelin-converting

 

enzyme (ECE-1)] producing the 21 amino acid residue active peptide. The

 

biochemistry      and    biology   of  the  endothelins     have    been    the  subject    of

 

several reviews.

 

PRINCIPLE OF THE ASSAY

 

The     microtiter   plate  provided     in  this  kit  has  been   pre-coated      with   an

 

antibody   specific   to   ET-1.   Standards   or   samples   are   then   added   to   the

 

appropriate      microtiter    plate   wells   with   a   biotin-conjugated       polyclonal

 

antibody preparation

 

                                              2


----------------------- Page 3-----------------------

specific for ET-1. and Avidin conjugated to Horseradish Peroxidase (HRP)

 

is   added   to   each   microplate   well   and   incubated.   Then   a   TMB   (3,3’5,   5’

 

tetramethyl-benzidine) substrate solution is added to each well. Only those

 

wells that contain ET-1, biotin-conjugated antibody and enzyme-conjugated

 

Avidin   will   exhibit   a   change   in   color.   The   enzyme-substrate   reaction   is

 

terminated by the addition of a sulphuric acid solution and the color change

 

is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The

 

concentration of ET-1. in the samples is then determined by comparing the

 

O.D. of the samples to the standard curve.

 

DETECTION RANGE

 

1.6    pg/ml-100    pg/ml.   The   standard    curve   concentrations    used    for  the

 

ELISA’s   were   100   pg/ml,   50   pg/ml,   25   pg/ml,   12.5   pg/ml,   6.2   pg/ml,   3.2

 

pg/ml,1.6pg/ml.

 

SPECIFICITY

 

This    assay    recognizes     recombinant      and   natural    mouse     ET-1     No

 

significant cross-reactivity or interference was observed.

 

SENSITIVITY

 

The   minimum   detectable   dose   of   mouse   ET-1   is   typically   less   than   0.8

 

pg/ml.

 

The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined

 

as the lowest protein concentration that could be differentiated from zero.

 

                                           3


----------------------- Page 4-----------------------

MATERIALS PROVIDED

 

               Reagent                                      Quantity

 

               Assay plate                                       1

 

               Standard                                          2

 

               Sample Diluent                               1 x 20 ml

 

               Biotin-antibody Diluent                      1 x 10 ml

 

               HRP-avidin Diluent                           1 x 10 ml

 

               Biotin-antibody                              1 x 120μl

 

               HRP-avidin                                   1 x 120μl

 

                                                            1 x 20 ml

               Wash Buffer

                                                         (25×concentrate)

 

               TMB Substrate                                1 x 10 ml

 

               Stop Solution                                1 x 10 ml

 

STORAGE

 

1.   Unopened   test   kits   should   be   stored   at   2-8°C   upon   receipt   and   the

 

    microtiter plate should be kept in a sealed bag. The test kit may be used

 

    throughout     the  expiration   date  of  the  kit,  provided   it  is  stored  as

 

    prescribed above. Refer to the package label for the expiration date.

 

2.   Opened test plate should be stored at 2-8°C in the aluminum foil bag

 

    with desiccants to minimize exposure to damp air. The kits will remain

 

    stable until the expiring date shown, provided it is stored as prescribed

 

    above.

 

3.   A   microtiter   plate   reader   with   a   bandwidth   of   10   nm   or   less   and   an

 

    optical   density   range   of   0-3   OD   or   greater   at   450nm   wavelength   is

 

    acceptable for use in absorbance measurement.

 

                                          4


----------------------- Page 5-----------------------

REAGENT PREPARATION

 

Bring all reagents to room temperature before use.

 

1.   Wash Buffer        If crystals have formed in the concentrate, warm up to

 

     room   temperature   and   mix   gently   until   the   crystals  have   completely

 

     dissolved. Dilute 20 ml of Wash Buffer Concentrate  into deionized or

 

     distilled water to prepare 500 ml of Wash Buffer.

 

2.   Standard       Centrifuge   the   standard   vial   at   6000-10000rpm   for 30s.

 

     Reconstitute     the Standard      with   1.0  ml   of Sample      Diluent.   This

 

     reconstitution    produces     a  stock   solution  of  100   pg/ml.   Allow   the

 

     standard to sit for a minimum of 15 minutes with gentle agitation prior to

 

     making   serial   dilutions.  The   undiluted   standard    serves   as  the  high

 

     standard (100 pg/ml). The Sample Diluent serves as the zero standard

 

     (0 pg/ml). Prepare fresh for each assay. Use within 4 hours and discard

 

     after use.

 

3.   Biotin-antibody        Centrifuge   the  vial  before   opening.   Dilute  to  the

 

     working      concentration      using    Biotin-antibody         Diluent(1:100),

 

     respectively.

 

4.   HRP-avidin       Centrifuge the vial before opening. Dilute to the working

 

     concentration using HRP-avidin Diluent(1:100), respectively.

 

Precaution: The Stop Solution provided with this kit is an acid solution. Wear

 

            eye, hand, face, and clothing protection when using this material.

 

OTHER SUPPLIES REQUIRED

 

      Microplate   reader   capable   of   measuring   absorbance  at   450   nm,   with

 

     the correction wavelength set at 540 nm or 570 nm.

 

                                           5


----------------------- Page 6-----------------------

      Pipettes and pipette tips.

 

      Deionized or distilled water.

 

      Squirt bottle, manifold dispenser, or automated microplate washer.

 

      An   incubator   which    can   provide   stable   incubation    conditions    up   to

 

     37°C±0.5°C.

 

SAMPLE COLLECTION AND STORAGE

 

     Serum      Use a serum separator tube (SST) and allow samples to clot

 

     for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove

 

     serum and assay immediately or aliquot and store samples at -20°C.

 

     Centrifuge   the   sample   again   after   thawing   before   the   assay.   Avoid

 

     repeated freeze-thaw cycles.

 

     Plasma       Collect    plasma     using    citrate,  EDTA,     or  heparin    as   an

 

     anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of

 

     collection.   Assay   immediately   or   aliquot   and   store  samples   at   -20°C.

 

     Centrifuge   the   sample   again   after   thawing   before   the   assay.   Avoid

 

     repeated freeze-thaw cycles.

 

Note: Grossly hemolyzed samples are not suitable for use in this as

 

ASSAY PROCEDURE

 

Bring    all  reagents    and   samples    to  room    temperature    before    use.   It  is

 

recommended that all samples, standards, and controls be assayed in duplicate.

 

All   the   reagents   should   be   added   directly   to   the   liquid   level   in   the   well.   The

 

pipette should avoid contacting the inner wall of the well.

 

                                             6


----------------------- Page 7-----------------------

1.    Add   100μl   of   Standard,   Blank,   or   Sample   per   well.   Cover   with   the

 

     adhesive strip. Incubate for 2 hours at 37°C.

 

2.    Remove the liquid of each well, don’t wash.

 

3.    Add 100μl of Biotin-antibody working solution to each well. Incubate

 

     for   1  hour   at  37°C.   Biotin-antibody   working          solution   may    appear

 

     cloudy.   Warm   up   to   room   temperature   and   mix   gently  until   solution

 

     appears uniform.

 

4.    Aspirate each well and wash, repeating the process three times for a

 

     total of three washes. Wash: Fill each well with Wash Buffer (200μl) and

 

     let   it  stand for  2  minutes, then   remove   the   liquid  by flicking  the  plate

 

     over a sink. The remaining drops are removed by patting the plate on a

 

     paper   towel.   Complete   removal   of   liquid   at   each   step   is   essential   to

 

     good performance.

 

5.    Add   100μl   of  HRP-avidin   working   solution   to   each   well.   Cover   the

 

     microtiter plate with a new adhesive strip. Incubate for 1 hour at 37°C.

 

6.    Repeat the aspiration and wash three times as step 4.

 

7.    Add 90μl of TMB Substrate to each well. Incubate for 10-30 minutes at

 

     37°C.    Keeping     the   plate   away    from    drafts   and   other   temperature

 

     fluctuations in the dark.

 

8.    Add    50μl   of  Stop   Solution   to    each    well  when    the   first   four  wells

 

     containing the highest concentration of standards develop obvious blue

 

     color. If color change does not appear uniform, gently tap the plate to

 

     ensure thorough mixing.

 

9.    Determine the optical density of each well within 30 minutes, using a

 

     microplate reader set to 450 nm.

 

                                              7


----------------------- Page 8-----------------------

CALCULATION OF RESULTS

 

Using   the   professional   soft   "Curve   Exert   1.3"   to   make   a   standard   curve   is

 

recommended, which can be downloaded from our web.

 

Average the duplicate readings for each standard, control, and sample and

 

subtract the average zero standard optical density. Create a standard curve

 

by reducing the data using computer software capable of generating a four

 

parameter   logistic   (4-PL)curve-fit.   As   an   alternative,   construct   a   standard

 

curve   by   plotting   the   mean   absorbance   for   each   standard   on   the   y-axis

 

against the concentration on the x-axis and draw a best fit curve through the

 

points on the graph. The data may be linearized by plotting the log of the

 

ET-1. concentrations versus the log of the O.D. and the best fit line can be

 

determined       by   regression     analysis.   This    procedure     will  produce     an

 

adequate but less precise fit of the data. If samples have been diluted, the

 

concentration      read   from   the  standard    curve    must   be  multiplied    by  the

 

dilution factor.

 

LIMITATIONS OF THE PROCEDURE

 

      The kit should not be used beyond the expiration date on the kit label.

 

      Do not mix or substitute reagents with those from other lots or sources.

 

      It is important that the Standard Diluent selected for the standard curve

 

     be consistent with the samples being assayed.

 

      If samples generate values higher than the highest standard, dilute the

 

     samples with the appropriate Standard Diluent and repeat the assay.

 

                                             8


----------------------- Page 9-----------------------

      Any    variation   in   Standard     Diluent,    operator,    pipetting    technique,

 

     washing   technique,   incubation   time   or   temperature,  and   kit   age   can

 

     cause variation in binding.

 

      This assay is designed to eliminate interference by soluble receptors,

 

     binding proteins, and other factors present in biological samples. Until

 

     all  factors    have   been    tested    in  the  Quantikine     Immunoassay,        the

 

     possibility of interference cannot be excluded.

 

TECHNICAL HINTS

 

      Centrifuge vials before opening to collect contents.

 

      When mixing or reconstituting protein solutions, always avoid foaming.

 

      To avoid cross-contamination, change pipette tips between additions of

 

     each standard level, between sample additions, and between reagent

 

     additions. Also, use separate reservoirs for each reagent.

 

      When   using   an   automated   plate   washer,   adding   a   30  second   soak

 

     period   following   the   addition   of   wash   buffer,   and/or   rotating   the   plate

 

     180 degrees between wash steps may improve assay precision.

 

      To   ensure   accurate   results,   proper   adhesion   of   plate   sealers   during

 

     incubation steps is necessary.

 

      Substrate Solution should remain colorless or light blue until added to

 

     the   plate.   Keep    Substrate     Solution    protected    from   light.  Substrate

 

     Solution   should   change   from   colorless   or   light   blue   to   gradations   of

 

     blue.

 

      Stop Solution should be added to the plate in the  same order as the

 

     Substrate Solution. The color developed in the wells will turn from blue

 

     to   yellow   upon   addition   of   the   Stop   Solution. Wells  that   are   green   in

 

     color indicate that the Stop Solution has not mixed thoroughly with the

 

     Substrate Solution.

 

                                              9

 