----------------------- Page 1-----------------------

     Human Vascuolar cell adhesion

                 molecule 1,VCAM-1

                            ELISA kit

 

                       Catalog No. CSB-E04753h

 

                                    (96 T)

 

    This immunoassay kit allows for the in vitro quantitative determination of human

 

    VCAM-1 concentrations in serum, plasma and other biological fluids.

 

    Expiration date   six months from the date of manufacture

 

    FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.

 

                                        1


----------------------- Page 2-----------------------

INTRODUCTION

 

VCAM-1   (vascular   cell   adhesion   molecule-1),   also   known   as

 

CD106,       is  the   name     of  both    a   human     gene    and   its   protein

 

product   which   have   a   considerable   role   in   the   human  immune

 

system.

 

The   VCAM-1   protein   mediates   the   adhesion   of   lymphocytes,

 

monocytes, eosinophils, and basophils to vascular endothelium.

 

It also functions in leukocyte-endothelial cell signal transduction,

 

and it may play a role in the development of atherosclerosis and

 

rheumatoid arthritis.

 

Upregulation of VCAM-1 in endothelial cells by cytokines occurs

 

as a result of increased gene transcription (e.g., in response to

 

Tumor   necrosis   factor-alpha   (TNF-α)   and   Interleukin-1   (IL-1))

 

and     through     stabilization     of  Messenger        RNA     (mRNA)       (e.g.,

 

Interleukin-4  (IL-4)).   The  promoter  region   of the  VCAM-1  gene

 

                                           2


----------------------- Page 3-----------------------

contains functional tandem NF-κB (nuclear factor-kappa B) sites.

 

The sustained expression of VCAM-1 lasts over 24 hours.

 

Primarily, the VCAM-1 protein is an endothelial ligand for VLA-4

 

(Very Late Antigen-4 or α4β1) of the β1 subfamily of integrins,

 

and     for  integrin   α4β7.     VCAM-1       expression      has   also   been

 

observed in other cell types (e.g., smooth muscle cells). It has

 

also been shown to interact with EZR and Moesin.

 

PRINCIPLE OF THE ASSAY

 

The microtiter plate provided in this kit has been pre-coated with

 

an antibody specific to VCAM-1. Standards or samples are then

 

added       to   the    appropriate      microtiter    plate    wells    with    a

 

biotin-conjugated antibody preparation specific for VCAM-1 and

 

Avidin conjugated to Horseradish Peroxidase (HRP) is added to

 

each     microplate     well   and   incubated.     Then     a  TMB     (3,3’,5,5’

 

tetramethyl-benzidine) substrate solution is added to each well.

 

                                         3


----------------------- Page 4-----------------------

Only      those     wells    that   contain     VCAM-1,        biotin-conjugated

 

antibody and enzyme-conjugated Avidin will exhibit a change in

 

color.    The     enzyme-substrate         reaction    is   terminated     by    the

 

addition   of   a   sulphuric   acid   solution   and   the   color   change   is

 

measured spectrophotometrically at a wavelength of 450 nm ± 2

 

nm.     The    concentration       of  VCAM-1       in  the   samples      is  then

 

determined        by   comparing       the   O.D.    of  the   samples      to  the

 

standard curve.

 

DETECTION RANGE

 

15.6 ng/ml-1000 ng/ml. The standard curve concentrations used

 

for   the   ELISA’s   were   1000   ng/ml,   500   ng/ml,   250   ng/ml,   125

 

ng/ml, 62.5 ng/ml, 31.2 ng/ml, 15.6 ng/ml.

 

SPECIFICITY

 

This assay recognizes recombinant and natural human VCAM-1.

 

No significant cross-reactivity or interference was observed.

 

                                          4


----------------------- Page 5-----------------------

SENSITIVITY

 

The minimum detectable dose of human VCAM-1 is typically less

 

than 3.9 ng/ml.

 

The sensitivity of this assay, or Lower Limit of Detection (LLD)

 

was   defined   as   the   lowest   protein   concentration   that   could   be

 

differentiated from zero.

 

MATERIALS PROVIDED

 

              Reagent                                 Quantity

 

             Assay plate                                   1

 

             Standard                                      2

 

             Sample Diluent                           1 x 20 ml

 

              Biotin-antibody Diluent                 1 x 10 ml

 

              HRP-avidin Diluent                      1 x 10 ml

 

              Biotin-antibody                         1 x 120l

 

              HRP-avidin                              1 x 120l

 

                                                      1 x 20 ml

             Wash Buffer

                                                   (25×concentrate)

 

             TMB Substrate                            1 x 10 ml

 

             Stop Solution                            1 x 10 ml

 

                                       5


----------------------- Page 6-----------------------

STORAGE

 

1.   Unopened   test   kits   should   be   stored   at   2-8°C   upon   receipt

 

    and the microtiter plate should be kept in a sealed bag. The

 

    test kit may be used throughout the expiration date of the kit,

 

    provided      it  is  stored    as   prescribed      above.    Refer    to   the

 

    package label for the expiration date.

 

2.   Opened test plate should be stored at 2-8°C in the aluminum

 

    foil bag with desiccants to minimize exposure to damp air. The

 

    kits will remain stable until the expiring date shown, provided it

 

    is stored as prescribed above.

 

3.   A  microtiter   plate  reader   with   a   bandwidth   of   10  nm  or less

 

    and an optical density range of 0-3 OD or greater at 450nm

 

    wavelength         is    acceptable        for     use     in    absorbance

 

    measurement.

 

REAGENT PREPARATION

 

Bring all reagents to room temperature before use.

 

1.  Wash   Buffer         If   crystals   have   formed   in   the   concentrate,

 

    warm      up    to  room     temperature      and    mix    gently   until   the

 

    crystals   have   completely   dissolved.   Dilute   20   ml   of   Wash

 

     Buffer Concentrate into deionized or distilled water to prepare

 

    500 ml of Wash Buffer.

 

                                          6


----------------------- Page 7-----------------------

2.  Standard         Centrifuge   the   standard   vial   at   6000-10000rpm

 

    for   30s.   Reconstitute   the  Standard  with   1.0   ml   of  Sample

 

     Diluent. This reconstitution produces a stock solution of 1000

 

     ng/ml. Allow the standard to sit for a minimum of 15 minutes

 

    with    gentle    agitation    prior  to   making     serial  dilutions.    The

 

     undiluted standard serves as the high standard (1000 ng/ml).

 

    The Sample Diluent serves as the zero standard (0 ng/ml).

 

     Prepare fresh for each assay. Use within 4 hours and discard

 

    after use.

 

3.   Biotin-antibody          Centrifuge the vial before opening. Dilute

 

    to     the     working      concentration        using    Biotin-antibody

 

     Diluent(1:100), respectively.

 

4.   HRP-avidin         Centrifuge the vial before opening. Dilute to the

 

    working       concentration       using   HRP-avidin         Diluent(1:100),

 

     respectively.

 

Precaution: The Stop Solution provided with this kit is an acid solution. Wear

 

            eye, hand, face, and clothing protection when using this material.

 

OTHER SUPPLIES REQUIRED

 

   Microplate reader capable of measuring absorbance  at 450

 

     nm, with the correction wavelength set at 540 nm or 570 nm.

 

                                          7


----------------------- Page 8-----------------------

   Pipettes and pipette tips.

 

   Deionized or distilled water.

 

   Squirt   bottle,   manifold   dispenser,   or   automated   microplate

 

    washer.

 

   An incubator which can provide stable incubation conditions

 

     up to 37°C ±0.5°C.

 

SAMPLE COLLECTION AND STORAGE

 

    Serum         Use     a   serum     separator     tube    (SST)     and    allow

 

    samples   to   clot   for   30   minutes   before   centrifugation   for   15

 

     minutes at 1000 g. Remove serum and assay immediately or

 

    aliquot   and   store   samples   at   -20°C.   Centrifuge   the       sample

 

    again      after   thawing      before     the   assay.     Avoid     repeated

 

    freeze-thaw cycles.

 

     Plasma       Collect plasma using citrate, EDTA, or heparin as

 

    an anticoagulant. Centrifuge for 15 minutes at 1000 g within

 

    30   minutes   of   collection.   Assay   immediately   or   aliquot   and

 

    store   samples   at   -20°C.   Centrifuge   the   sample   again           after

 

    thawing before the assay. Avoid repeated freeze-thaw cycles.

 

Note: Grossly hemolyzed samples are not suitable for use in this assay.

 

                                          8


----------------------- Page 9-----------------------

ASSAY PROCEDURE

 

Bring    all  reagents   and   samples    to  room   temperature   before    use.   It  is

 

recommended that all samples, standards, and controls be assayed in duplicate.

 

All   the   reagents   should   be   added   directly   to   the   liquid   level   in   the   well.   The

 

pipette should avoid contacting the inner wall of the well.

 

1.    Add 100l of Standard, Blank, or Sample per well. Cover with

 

     the adhesive strip. Incubate for 2 hours at 37°C.

 

2.    Remove the liquid of each well, don’t wash.

 

3.    Add 100l of Biotin-antibody working solution to each well.

 

     Incubate       for   1   hour     at  37°C.    Biotin-antibody           working

 

     solution may appear cloudy. Warm up to room temperature

 

     and mix gently until solution appears uniform.

 

4.    Aspirate   each   well   and   wash,   repeating   the   process   three

 

     times for a total of three washes. Wash: Fill each well with

 

     Wash   Buffer   (200l)          and   let  it  stand   for   2   minutes,   then

 

     remove       the   liquid   by   flicking   the    plate   over    a  sink.    The

 

     remaining drops are removed by patting the plate on a paper

 

     towel. Complete removal of liquid at each step is essential to

 

     good performance.

 

                                            9


----------------------- Page 10-----------------------

 5.  Add    100l     of HRP-avidin   working         solution    to  each    well.

 

    Cover the microtiter plate with a new adhesive strip. Incubate

 

    for 1 hour at 37°C.

 

 6.  Repeat the aspiration and wash five times as step 4.

 

 7.  Add 90l of TMB Substrate to each well. Incubate for 10-30

 

    minutes   at   37°C.   Keeping   the   plate   away   from   drafts       and

 

    other temperature fluctuations in the dark.

 

 8.  Add 50l of Stop Solution to each well   when the first four

 

    wells     containing     the   highest    concentration       of  standards

 

    develop obvious blue color. If color change does not appear

 

    uniform, gently tap the plate to ensure thorough mixing.

 

 9.  Determine the optical density of each well within 30 minutes,

 

    using a microplate reader set to 450 nm.

 

CALCULATION OF RESULTS

 

Using   the   professional   soft   "Curve   Exert   1.3"   to   make   a   standard   curve   is

 

recommended, which can be downloaded from our web.

 

Average the duplicate readings for each standard, control, and

 

sample and subtract the average zero standard optical density.

 

Create   a   standard   curve   by   reducing   the   data   using computer

 

software capable of generating a four parameter logistic (4-PL)

 

                                        10


----------------------- Page 11-----------------------

curve-fit. As an alternative, construct a standard curve by plotting

 

the   mean   absorbance   for   each   standard   on   the   y-axis   against

 

the concentration on the x-axis and draw a best fit curve through

 

the points on the graph. The data may be linearized by plotting

 

the log of the VCAM-1 concentrations versus the log of the O.D.

 

and the best fit line can be determined by regression analysis.

 

This procedure will produce an adequate but less precise fit of

 

the data. If samples have been diluted, the concentration read

 

from the standard curve must be multiplied by the dilution factor.

 

LIMITATIONS OF THE PROCEDURE

 

   The kit should not be used beyond the expiration date on the

 

    kit label.

 

   Do not mix or substitute reagents with those from other lots or

 

    sources.

 

   It   is  important    that  the   Standard     Diluent   selected   for   the

 

    standard      curve     be   consistent     with   the   samples      being

 

    assayed.

 

   If samples generate values higher than the highest standard,

 

    dilute the samples with the appropriate Standard Diluent and

 

    repeat the assay.

 

                                       11


----------------------- Page 12-----------------------

   Any       variation    in   Standard      Diluent,    operator,     pipetting

 

    technique,        washing       technique,       incubation       time     or

 

    temperature, and kit age can cause variation in binding.

 

   This   assay   is   designed   to   eliminate   interference   by   soluble

 

    receptors,      binding    proteins,   and    other   factors    present    in

 

    biological samples. Until all factors have been tested in the

 

    Quantikine       Immunoassay,         the   possibility   of   interference

 

    cannot be excluded.

 

TECHNICAL HINTS

 

   Centrifuge vials before opening to collect contents.

 

   When mixing or reconstituting protein solutions, always avoid

 

    foaming.

 

   To   avoid   cross-contamination,   change   pipette   tips  between

 

    additions of each standard level, between sample additions,

 

    and between reagent additions. Also, use separate reservoirs

 

    for each reagent.

 

   When using an automated plate washer, adding a 30 second

 

    soak     period   following    the   addition   of  wash    buffer,  and/or

 

    rotating   the   plate   180    degrees     between     wash   steps   may

 

    improve assay precision.

 

                                        12


----------------------- Page 13-----------------------

   To ensure accurate results, proper adhesion of plate sealers

 

    during incubation steps is necessary.

 

   Substrate Solution should remain colorless or light blue until

 

    added to the plate. Keep Substrate Solution protected from

 

    light. Substrate Solution should change from colorless or light

 

    blue to gradations of blue.

 

   Stop Solution should be added to the plate in the same order

 

    as the Substrate Solution. The color developed in the wells

 

    will turn from blue to yellow upon addition of the Stop Solution.

 

    Wells that are green in color indicate that the Stop Solution

 

    has not mixed thoroughly with the Substrate Solution.

 

                                        13


----------------------- Page 14-----------------------

Notes

 

                                                                            14


----------------------- Page 15-----------------------

                                                                            15


----------------------- Page 16-----------------------

                                                                            16


----------------------- Page 17-----------------------

                                                                            17