三聚氰胺分析 Rtx-5MS GC/MS SIM

GC_FF00978

1 氰尿酸 Cyanuric acid 108-80-5
2 三聚氰胺一酰胺 Ammelide 645-93-2
3 三聚氰胺二酰胺 Ammeline 645-92-1
4 三聚氰胺 Melamine 108-78-1
5 苯代三聚氰胺 Benzoguanamine 91-76-9



 

image
 
Column Rtx®-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Sample melamine, cyanuric acid, ammelide, ammeline, benzoguanamine
Conc.: 10 µg/mL prederivatized
Injection
Inj. Vol.: 1.0 µL splitless (hold 1 min)
Liner: 3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.: 280 °C
Oven
Oven Temp: 75 °C to 320 °C at 15 °C/min (hold 4 min)
Carrier Gas He, constant flow
Flow Rate: 1 mL/min
Detector MS
Mode: SIM
Transfer Line Temp.: 290 °C
Ionization Mode: EI
 



 

推荐仪器
  • *
  • *
  • *
  • *